dc.contributor.author |
Ermir PAPADHIMITRI; Department of electronic, Informatics and statistics, University of Bologna |
|
dc.contributor.author |
Nicola TESTONI; Department of electronic, Informatics and statistics, University of Bologna |
|
dc.contributor.author |
Luca DE MARCHI; Department of electronic, Informatics and statistics, University of Bologna |
|
dc.contributor.author |
Guido MASETTI; Department of electronic, Informatics and statistics, University of Bologna |
|
dc.date |
2013-06-18 05:59:22 |
|
dc.date.accessioned |
2013-07-15T11:52:05Z |
|
dc.date.accessioned |
2015-11-23T16:01:21Z |
|
dc.date.available |
2013-07-15T11:52:05Z |
|
dc.date.available |
2015-11-23T16:01:21Z |
|
dc.date.issued |
2013-07-15 |
|
dc.identifier |
http://ecs.epoka.edu.al/index.php/iscim/iscim2011/paper/view/779 |
|
dc.identifier.uri |
http://dspace.epoka.edu.al/handle/1/750 |
|
dc.description.abstract |
It is well known that ultrasonography is a diagnostic method for visualizinginside human tissues by spreading ultrasounds and measuring their return time tothe sensor. However, the interface between the human skin and this ultrasoundtransducer attenuates the received signal and the medical image quality deterioratessignificantly. In this paper we propose a filtering technique in order to compensatethis attenuation. A finite impulse response filter (FIR) based on a Delayed LeastMean Square (DLMS) was optimized and implemented. The main contribution ofour work consists of finding the order and the coefficients of the filter that minimizethe attenuation error. We validate our method first on simulated data and later on areprogrammable FPGA device for a real time performance testing. Among others,we show that incrementing the order of the filter, not always is the best way toreduce image quality errors. |
|
dc.format |
application/pdf |
|
dc.language |
en |
|
dc.publisher |
International Symposium on Computing in Informatics and Mathematics |
|
dc.source |
International Symposium on Computing in Informatics and Mathematics; 1st International Symposium on Computing in Informatics and Mathematics |
|
dc.title |
A Filtering technique based on a DLMS algorithm for ultrasonography video |
|
dc.type |
Peer-reviewed Paper |
|